Technical Indicators | Details |
---|---|
Thermocouple Types | K.J.T.E.R.S.N Types |
Measurement Range | K-type Thermocouple: -200.0°C~1372°C (-328.0°F~2501°F) J-type Thermocouple: -210.0°C~1200°C (-346.0°F~2192°F) T-type Thermocouple: -250.0°C~400.0°C (-418.0°F~752.0°F) E-type Thermocouple: -150.0°C~1000°C (-238.0°F~1832°F) R-type Thermocouple: 0°C~1767°C (32°F~3212°F) S-type Thermocouple: 0°C~1767°C (32°F~3212°F) N-type Thermocouple: -200.0°C~1300°C (-328.0°F~2372°F) |
Measurement Accuracy | ±(0.2% + 0.5°C) or ±(0.2% + 0.9°F) |
Characteristics | |
Display Resolution | T < 1000: 0.1°C/ °F / K T≥1000: 1.0°C/ °F / K |
Repeatability Accuracy | ±0.1% + 0.3°C / ±0.1% + 0.5°F |
Temperature Coefficient | When the ambient temperature is outside the specified temperature range of 18°C to 28°C: 0.01% of the reading + 0.05°C/°C (0.05°F/ °F) When measuring temperatures below -100°C, add 0.1% of the reading |
Response Time | Approximately 500ms |
Items | Details |
---|---|
Supported Functions | Unit switching among ºC, ºF, K Data retention Battery level display Difference Timed measurement Data recording (72000 sets) Type-C USB, Bluetooth data transmission PC software Mobile APP |
Battery Type | 3*AAA alkaline batteries |
Anti-power Frequency Interference | 50/60Hz |
Operating Environment Temperature | -10°C~50°C (14°F~122°F) |
Operating Environment Humidity | <90%RH (non-condensing) |
Storage Environment Temperature | -20°C~60.0°C (-4°F~140°F) |
Drop Test | 1m |
EMC Standards | RE: 30MHz - 1GHz CLASS A standard EN 61326-1:2013 Reference EN55011:2009+A1:2010 RI: 80MHz - 1GHz 3V/m; 1.4GHz - 6GHz 3V/m CLASS A standard EN 61326-1:2021 Reference EN61000-4-3:2006+A1:2008+A2:2010 ESD: Contact 4kv Air 8kv CLASS B standard EN 61326-1:2013 Reference EN61000-4-2:2009 |
Resound Research Institute continuously introduces advanced new technologies and products in industrial application scenarios such as market supervision, public safety, environmental monitoring, and science, education, research, and production. This has led to the formation of a domestically produced high-end scientific equipment industry base that integrates design, research and